Operating Ranges |
0 to .5 mR/hr; 0 to 5 mR/hr; 0 to 50 mR/hr;
0 to 300 CPM; 0 to 3000 CPM; 0 to 30000 CPM |
|
Accuracy |
± 20% |
|
Linearity |
± 20% |
|
Warm Up Time |
None |
|
Response Time |
5 second time constant |
|
Temperature Range |
- 10 to + 50° C (14° to 122° F) |
Relative Humidity Rng |
20% to 95% |
|
Visual Readout |
Easy-to-read 2.62" round meter |
|
Audio Readout |
Speaker; clicking type headset is optional |
|
Detector Voltage |
920 volts |
|
Controls |
4 position selector switch (off; x1; x10; x100); Battery check &
Speaker on/off switch |
|
Power |
One 9 volt cell for speaker (NEDA 1604-A or equal) |
|
Batteries |
2 standard “D” cells (1.5 volt) |
|
Connectors |
Microphone connection for optional headset; BNC for probe connection |
| Calibration |
Pre-calibrated at the factory |
|
Case |
Ruggedized Aluminum body and cover |
|
Finish |
Polyurethane enamel |
|
Dimensions |
4"W x 9"L x 5" H.
|
|
Weight |
4.2 lbs. including detector |
|
Warranty |
1 year on meter (see probe for separate warranty) |
| Radiation Detected |
Beta, Gamma, X-ray |
|
Background |
<15CPM |
|
Temperature Range |
- 55° to + 75° C (- 67° to + 167° F) |
|
Wall thickness of detector |
30-45mg/cm² |
|
Wall Material |
446 Stainless Steel |
|
Quench Gas |
Halogen |
|
Active Length |
0.75" |
|
Diameter of Detector (effective) |
0.625" |
|
Wall Thickness of Sliding Window |
3.64 g/cm2 |
|
Window Material |
Brass |
|
Window Size (total area) |
0.5" x 0.9" (1.3cm x 0.8cm) |
|
Detector Size |
5.1" x .94" (13cm x 2.4cm) |
|
Weight of Detector w/Cable & Connector |
12 oz. (340g) |
|
Cable Length (Straight) |
3 ft. |
|
Operating Voltage |
900V |
|
Nominal Charge per Pulse: |
2.7 x 10-9 coulombs |
|
Warranty |
One yr on probe, 6 mo. on GM tube |